|
Anions
1. Materials: 1. Dionex chromatographer 2. stock solutions for Cl-1 and SO42 3. 7 100-ml volumetric flasks 4. Nanopure water 5. 100- and 1000-l pipettes (calibrated), tips 6. 7 125 ml poly bottles to store standards
2. If Total Anion concentration is < 600 um the Dionex can be set up to run on the Fast column (~ 3 min / sample). Make sure that the Dionex has plenty of regenerant and eluent.
Regenerant (2 liters)
100 ml .5 N H2SO4 (27.7 ml H2SO4 in 2 liter DI) Fill to 2 Liters with DI
Eluent (2 liters) Should be relatively fresh.
.4240g Na2CO3 and 0.0252 g NaHCO3 Fill to 2 liters
3. Standard stock solutions for cloride and sulfate can be made in the same volumetric
Cloride = Sodium Cloride NaCl (FW = 58.44) 10 mM NaCl = .05844 g / 100 ml
Sulfate = Sodium Sulfate Na2SO4 (FW = 142.04) 10 mM Na2SO4 = .142. g / 100ml
4. Standards: 1. 2 uM Cloride & sulfate = 20 ul 10 mM NaCl & 20 ul 10 mM Na2SO4 /100 ml DI 2. 5 uM Cloride & sulfate = 50 ul 10 mM NaCl & 50 ul 10 mM Na2SO4 /100 ml DI 3. 10 uM Cloride & sulfate = 100 ul 10 mM NaCl & 100 ul 10 mM Na2SO4 /100 ml DI 4. 15 uM Cloride & sulfate = 150 ul 10 mM NaCl & 150 ul 10 mM Na2SO4 /100 ml DI 5. 30 uM Cloride & sulfate = 300 ul 10 mM NaCl & 300 ul 10 mM Na2SO4 /100 ml DI
5. The attenuation will need to be changed for the different concentrations in the standard curve. On the fast column the following work well. For actual area under the curve multiply the attenuation by the area.
6. To run a sample: 1. Inject sample into port (>1 ml) 2. Hit [Load/ Inject] (valve will switch) 3. Change attenuation if needed 4. Plot sample [Shift/Down] [Plot] [Enter]
|
||||||||||||||||||||
|
|
Please contact arc_im@mbl.eduwith questions, comments, or for technical assistance regarding this web site. | |||||||||||||||||||