Dissolved Methane Probe Information

Information on the experimental dissolved methane probe construction and calibration is presented here.  Note, as currently configured, this probe as several limiations; namely,
The dCH4 probe sensor is a Figaro TGS2611-C00 metal oxide semiconductor whose resistence decreases as a function of increasing methane concentration. By use of a simple voltage divider, methane concentration can be estimated, as given by this circuit that applies 5V across both the chip heater (RH) and the sensor  (RS).
Sensor Circuit

Voltage drop across the load resistor  (RL variable resistor adjusted between 5k and 8k ohms) is monitored with a Weeder Technologies WTAIN-M RS232 A2D board. Relationship between voltage, VL, and dissolved methane is obtained via calibration with methane standards (see below).

The Figaro sensor is packaged into a 1/2" polycarbonate tube that has Luer-Lok fittings that connect to a 4" piece of silicone tubing.  An o-ring sits in a 1/16" grove cut into the polycarbonate tube, which holds the senor in place by friction.  A GL45 cap with a modified nylon 1/2" compression union fitting allows the probe to be appropriately placed in the microcosm. An image of the finished probe is shown here:

dCH4 probe housing

Methane and oxygen diffuse across the silicone tubing and adsorb to the semiconductor in the sensor.

Dissolved methane probes were calibrated by placing probes in micrcosm reactors filled with DI water and sparged with mixtures of methane and air. Concentration of methane in gas phase was measured in a manner identical to that in the main experiment.  Because temperature and transport characteristics strongely influence probe response, conditions for probe calibration were set as close as possible to those found in the experimental system.

Probe voltage output as a function of methane gas cocnetration is show here:
Voltage output
Note, probe response delays are due to transport of methane across silicone tubing and time required to flush reactor headspace with desired methane concentraion.  All four probes were in the same reactor, which contained approximately 3 L of DI water and 3 L of gas headspace, and gas flow rates varied from 35  to 110 mL min-1.  

The probe volages were fit to the following logrithmic equation:

            v = vM Ln(PCH4 + kCH4) + v0

where PCH4 is the partial pressure of methane (%), and the three parameters were determined by least squares fit to the data presented above.  This equation fit the data fairly well as shown here:

Voltage vs CH4 concentration

Recalibration and Tubing Change

After several days, water accumulates in the silicone tubing, which block methane transport and causes voltage drops.  Larger diameter silicone tubing is being tested (3/16" OD, 1/8" ID, 2 3/4" long).  Also, resistence of the load resistors (RL) were decreased (to 2 - 3 kohms) and adjusted to that all probes have similar outputs at CH4 concentrations in equilibrium with 1.5% CH4 gas headspace.